Total Estrogen (E1/E2/E3) ELISA Kit |
|
Total Estrogen (E1/E2/E3) ELISA Kit
- ES monoclonal antibody binds exclusively with E1, E2 and E3, and
does not show cross-reaction with other chemicals of similar structures. A monoclonal
antibody is uniform in quality, generating very little lot-to-lot variation.
- The quantitative analysis ranges from 0.1g/L to 3g/L (ppb).
- The ELISA measurement is highly reproducible; the coefficient
of variation (CV) is mostly under 10%.
- The assay requires less amount of harmful solvent than instrument
analyses.
- With ease of handling, the total time for measurement is only
2.5 hours.
- The kit, a 96-well microplate format, enables simultaneous measurement
of multiple samples at more reasonable cost.
Chemical Structures of Targets

Standard Curves

Pretreatment Methods
See here.
Test Data
Recovery Test using River Water
(n=4) |
|
@ |
ES ELISA Kit |
Unspiked |
Determination
(ng/L) |
12.4 |
SD
(%) |
6.5 |
Spiked with 2ng/L E2 |
Determination
(ng/L) |
14.4 |
SD
(%) |
9.2 |
Recovery
(%) (n=4) |
100.0 |
|
Comparison with LC/MS/MS
Comparison with LC/MS/MS |
|
@ |
E1+E2 (ng/L) |
Sample |
LC/MS/MS |
ES ELISA Kit |
A |
2.08 |
3.75(1.8) |
B |
4.25 |
5.36(1.3) |
|
|
Figures in parentheses indicate relative value to LC/MS/MS data. |
Cross-reactivity Pattern
Estrogens |
|
Compound |
% Reactivity |
ES ELISA Kit |
Estron (E1) |
87.0 |
2-methoxy E1 |
<0.03 |
E1-3-sulfate |
<0.03 |
17-Estradiol (E2) |
100.0 |
16-keto E2 |
118.0 |
2-methoxy E2 |
0.2 |
E2-17-glucronide |
5.0 |
E2-3-glucronide |
<0.03 |
E2-3-sulfate-17-glucronide |
0.5 |
Ethynylestradiol |
0.7 |
Estriol (E3) |
55.0 |
16-epi-E3 |
129.0 |
E3-16-glucronide |
48.0 |
|
Other hormones |
|
Compound |
% Reactivity |
ES ELISA Kit |
cis -Androsterone |
<0.03 |
trans -Androsterone |
<0.03 |
Cholesterol |
<0.03 |
Dehydroisoandrosterone |
<0.03 |
5-Dihydroteststerone |
<0.03 |
Hydrocortisone |
<0.03 |
Pregnenolone |
<0.03 |
Teststerone |
<0.03 |
|
Kit Format
Kit Format |
Comment |
Microplate |
1 plate (96 wells)& reagents |
Needs a microplate reader (450nm)
For multiple and simultaneous measurement |
For research use only. Not for diagnostic use.
|